ICP Tests for Reef Aquariums: 8 Questions to Answer Before You Touch a Dosing Bottle

Four red flags, one email, and a shelf full of dosing bottles. Before you touch any of them, here is how to read an ICP test the way a chemist would: slowly, sceptically, and with natural seawater as the yardstick.

The lab email lands at 7 a.m. You open the PDF and the bars jump out at you: iodine low, nickel high, lithium “not detected”, and a zinc value printed in red. Your hand is already reaching for the trace-element bottles. Stop there for a moment. An ICP test is a powerful snapshot of your water, but it is a snapshot taken through a difficult lens, and most of the useful work happens before you dose anything. Reading the report properly comes first, and that is what this article is about.

Below are eight questions worth answering once the results arrive, each answered directly first and then explained.

62–661 ppbiodine reported by five labs for the same tank water [1]

0.2%maximum dissolved solids in EPA’s ICP methods; seawater is about 3.5% [4][5][15]

0.15 µg/kgmean copper in the natural ocean [8]

4 µg/Lcopper level that harmed two coral species in a 5-week study [11]

What does an ICP test actually measure, and what does it miss?

An ICP test measures how much of each chemical element is in the water, not which compound that element is in. It is excellent for metals and many trace elements, but it is not a replacement for your alkalinity, nitrate or phosphate kits.

EPA Method 200.7, the classic ICP-OES method for water, describes the technique as determining “metals and some nonmetals in solution” [4]. Its analyte list runs from aluminium to zinc and includes phosphorus and silica, while nitrogen and iodine are not on it [4]. The ICP-MS method, EPA 200.8, lists 21 elements and includes neither phosphorus nor iodine [5].

Three practical consequences follow:

  • Phosphate. ICP reports phosphorus; a phosphate figure on the report is calculated from that measured phosphorus [1]. Most hobby kits instead measure orthophosphate, which Hanna describes as the only form of phosphate that can be determined directly; a total-phosphorus test needs digestion first [16]. The two numbers are related but not the same measurement. For why phosphate matters at all, see our guide to phosphate in reef aquariums.
  • Alkalinity and nitrate. Alkalinity is a property of the carbonate system, not an element, and nitrogen is absent from the EPA ICP-OES analyte list [4]. Keep testing these at home; Paletta’s review stresses that getting the main parameters stable comes first [3]. Our overview of essential water parameters in a reef aquarium covers that routine.
  • Iodine. Iodine is notoriously awkward: an analytical paper on ICP-MS describes memory effects and recovery problems “mainly due to the easy volatilization of iodine” [14]. In the five-lab hobby comparison, iodine results ranged from 62 to 661 ppb, a more than tenfold spread [1].

How does ICP-OES work, and why is seawater such a hard sample?

The lab sprays your water into an argon plasma hot enough to break it into atoms and ions, then either reads the light they emit (OES) or weighs the ions (MS). Salt is the problem: seawater carries far more dissolved solids than these methods are designed for.

In ICP-OES, “samples are nebulized and the resulting aerosol is transported to the plasma torch”, where a radio-frequency inductively coupled plasma produces element-specific emission spectra; a spectrometer then measures the intensity of each element’s lines [4]. In ICP-MS, the plasma ionises the sample, the ions are pulled through a vacuum interface and “separated on the basis of their mass-to-charge ratio by a quadrupole mass spectrometer” [5]. Put simply: OES identifies elements by their colour of light, MS by their weight.

How an ICP test works: sample, nebuliser, argon plasma and an OES or MS detector
From vial to result: the main stages of an ICP analysis, by optical emission (OES) or mass spectrometry (MS), based on the EPA method summaries [4][5].

The salt problem behind every ICP test

Both EPA methods say dissolved solids should stay below 0.2% (w/v) [4][5]. Seawater at practical salinity 35 contains about 35.165 g of salt per kilogram [15], roughly 3.5%. On those numbers, a lab following that guideline would need to dilute your sample roughly 18-fold, and every trace element drops by the same factor. EPA notes that high salt causes “salt buildup at the tip of the nebulizer” and “analyte signal suppressions” [4]; Thermo Fisher’s seawater application note likewise calls salt-containing samples “a challenge for ICP-OES” [13].

The effect is measurable. In a 2023 study on tin, undiluted model seawater of only 6‰ and 18‰ salinity cut the tin signal by up to 40% on ICP-OES and up to 80% on ICP-MS, and with direct injection the limits of quantification rose with salinity [12]. ICP-MS has its own trap: chlorine forms ArCl⁺ at mass 75, overlapping arsenic, and ClO⁺ at mass 51, overlapping vanadium, so corrections are mandatory [5]. The EPA methods deal with this through dilution, internal standards and interference corrections [4][5]; the hobby comparisons cited here do not report how each lab handles it, which leads straight to the next question.

Why do two labs give different numbers for the same water?

Because, as the hobby comparisons point out, there is no verified seawater standard that covers all the trace elements labs report, and for many trace elements the labs do not agree [1][2].

In Sanjay Joshi’s 2023 ICP test comparison, water drawn from one 500-gallon reef within 20 minutes went to five labs [1]. Iodine came back between 62 and 661 ppb, lithium between 416 and 817 ppb at four labs while a fifth listed 0, and lanthanum between 0 and 17 ppb [1]. Each vendor also sets its own “acceptable” ranges, and those differ from vendor to vendor [1].

A 2024 follow-up by Joshi and Ross sent five replicate samples of the same water to each of three anonymous vendors [2]. Each vendor was repeatable for major elements, with coefficients of variation as low as 0.51% for chloride and 1.94% for calcium, but variation grew at low trace concentrations, and the vendors’ averages differed significantly for every element except bromine [2]. Neither study tried to decide which lab was accurate [1][2].

ElementLowest labHighest lab
Iodine62 ppb661 ppb
Lithium0 ppb817 ppb
Lanthanum0 ppb17 ppb
Nickel21 ppb38.41 ppb

Same water, five labs: selected results from Joshi’s 2023 review [1].

Certified seawater reference materials do exist, such as NRC’s NASS-7 for trace metals [17], but the hobby comparisons note that accuracy for reef-lab reports is still unresolved without a shared, verified reference [1][2]. Joshi and Ross suggest that keepers who track their tank over time stick with one vendor [2].

Your ICP test is most trustworthy when it is compared with itself: same lab, same method, over time.

What does “below detection limit” really mean?

It means the lab could not reliably tell your sample apart from a blank. It does not mean the element is absent.

The US EPA defines the method detection limit as the minimum measured concentration that can be reported “with 99% confidence that the measured concentration is distinguishable from method blank results” [7]. Anything below that line is simply too small to separate from background noise for that method, that day.

Iron is the classic example. The natural ocean averages about 0.03 µg/kg of iron [8], while one manufacturer’s ICP-OES setup, tested on simulated seawater, reported an iron detection limit of 7.5 µg/L [13]. A “not detected” iron result on an ICP test is therefore expected even in healthy water, and is not by itself a reason to dose. Our article on iron in marine aquariums explains why careful monitoring matters more than chasing a number. The lithium “0” in Joshi’s comparison, against 416 to 817 ppb elsewhere, shows that a zero on one report can sit next to large values on others [1].

Which red flags on an ICP test are often sampling artefacts?

High boron or silicon after using glass, and high zinc, copper or other metals after contact with metal, gloves, dust or unrinsed containers are the usual suspects. Low trace levels can also change during shipping.

At concentrations this low, contamination is easy. EPA Method 1669 warns that much historical trace-metal data in natural waters was “erroneously high” and lists metal equipment, containers, dust, human contact and even talc gloves “that contain high levels of zinc” as sources [6]. Oceanographers face the same problem with steel wire, rust, dust and copper- and zinc-based antifouling paints [10]. EPA 200.7 adds that borosilicate glass can contaminate boron and silica results, so only plastic, PTFE or quartz labware should be used [4].

ICP test sampling do's and don'ts: use the lab vial, avoid glass, metal and dust
Sampling habits that keep contamination out of your ICP test vial [1][3][4][6].

Handling during transit matters too. Two of the five labs in Paletta’s comparison asked for filtered samples, so biological activity would not alter the water on its way to the lab, and only one lab compensated for samples not at 35 ppt salinity [3]. If your salinity is off, every element will read proportionally off as well.

Which results need action now, and which can wait for a retest?

A clearly raised copper result deserves a prompt confirmation and a plan. Most other single out-of-range trace values can wait for a second ICP test [1].

The myth

“Every red bar on the report is something to fix today.”

Joshi’s lab comparison concluded the opposite: ICP is most useful for spotting large excursions, less reliable for fine-tuning individual trace elements, and readers should “evaluate trends” rather than take the numbers as absolutes [1]. Paletta adds that there is still too little information about what each element does [3].

Copper is the exception worth taking seriously. The ocean mean is about 0.15 µg/kg [8]. In a five-week laboratory study, copper exposures from 2 to 20 µg/L were tested on three coral species, and effects on Acropora cervicornis and Pocillopora damicornis appeared at concentrations as low as 4 µg/L; the effects reported for these two species included reduced skeletal growth [11]. That gap is large enough that a copper flag should be confirmed, not ignored.

Natural seawater reference values for elements reported on an ICP test, from an oceanographic summary table
Mean concentrations in the natural ocean, a neutral yardstick for any ICP report [8][11].
ElementNatural ocean mean
Strontium7,800 µg/kg
Boron4,500 µg/kg
Lithium180 µg/kg
Iodine58 µg/kg
Barium15 µg/kg
Molybdenum10 µg/kg
Nickel0.48 µg/kg
Zinc0.35 µg/kg
Copper0.15 µg/kg
Iron0.03 µg/kg

Converted from the MBARI Periodic Table of Elements in the Ocean summary table (ng/kg ÷ 1,000), values mainly from Nozaki [9], as compiled by MBARI [8]. 1 µg/kg is roughly 1 ppb.

Remember that open-ocean trace metals vary with depth and region [10]; these are averages, not targets to dose to. In AVen’s view, a reef tank does not need to match the open ocean to the decimal, but a value tens of times above it is a fair trigger for a closer look.

How should I correct a real problem without making it worse?

Confirm the ICP test first, change slowly, and remove the cause rather than chasing the number.

Decision flowchart for a red flag on an ICP test in a reef aquarium
Red flag on your ICP test? A decision path built on the sources in this article [1][2][3][4][6][11].

Paletta’s advice is to adjust any trace element “only gradually”, over days or weeks, not overnight [3]. Stability of core parameters such as alkalinity matters more than any single ICP value [3]; keep your calcium, alkalinity and magnesium balance in order, as explained in the triad in a marine aquarium. When something is genuinely too high, dilution is the gentlest tool: our guide on the importance of water changes explains the routine.

Two of the labs in Joshi’s review also tested RO/DI source water, which he suggests can help trace where contamination comes from [1]. If a metal is high in both, the problem may be upstream of the tank.

What this means for your tank

  1. Check salinity and your home-kit values (alkalinity, nitrate, phosphate) before reading any trace bar [3].
  2. Compare every flag with natural seawater, not only with the lab’s own range [1][8].
  3. Treat “below detection” as unknown, not zero [7].
  4. Ask whether glass, metal, gloves or dust touched the sample [4][6].
  5. Retest with the same lab before dosing, and add an RO/DI sample if metals are high [1][2].
  6. If copper is confirmed, make it the priority: find the source and dilute with water changes (AVen’s opinion, given the coral effects reported at 4 µg/L [11]).
  7. Correct genuine shortfalls gradually, over days or weeks [3].

How often should I order an ICP test?

None of the sources cited here sets a fixed schedule. A baseline of a few tests while the tank is stable, then periodic checks with the same lab, is a sensible pattern.

Paletta ran one ICP test a month for six months to build a baseline, and advises new tanks to wait until they are stable before relying on ICP [3]. Paletta describes how Andrew Sandler of PoloReef uses regular ICP tests to establish a baseline “for when things are good”, so negative trends show up early [3]. Sticking with one vendor keeps those trends comparable [2].

My rule of thumb, as an opinion: one red bar is a question, two consecutive red bars from the same lab are a trend, and only a trend that agrees with what your corals show deserves a change in dosing. The next time that 7 a.m. email arrives, open the report, put the bottles down, and start with the first question in this list.

References

  1. Joshi, S. (2023). The Reef Builders big ICP test review by Sanjay Joshi. Reef Builders. https://reefbuilders.com/2023/07/12/the-reef-builders-big-icp-test-review-by-sanjay-joshi
  2. Joshi, S., & Ross, R. (2024). Evaluating the repeatability and differences between ICP tests. Reef Builders. https://reefbuilders.com/2024/03/06/evaluating-the-repeatability-and-differences-between-icp-tests
  3. Paletta, M. (2023). A beginner’s guide to ICP testing. Reef Builders. https://reefbuilders.com/2023/08/05/a-beginners-guide-to-icp-testing
  4. Martin, T. D., Brockhoff, C. A., Creed, J. T., & EMMC Methods Work Group (1994). Method 200.7, Revision 4.4: Determination of metals and trace elements in water and wastes by inductively coupled plasma-atomic emission spectrometry. U.S. EPA. https://epa.gov/sites/default/files/2015-08/documents/method_200-7_rev_4-4_1994.pdf
  5. Creed, J. T., Brockhoff, C. A., & Martin, T. D. (1994). Method 200.8, Revision 5.4: Determination of trace elements in waters and wastes by inductively coupled plasma-mass spectrometry. U.S. EPA. https://19january2017snapshot.epa.gov/sites/production/files/2015-06/documents/epa-200.8.pdf
  6. U.S. EPA (1996). Method 1669: Sampling ambient water for determination of metals at EPA water quality criteria levels. U.S. EPA. https://19january2021snapshot.epa.gov/sites/static/files/2015-10/documents/method_1669_1996.pdf
  7. U.S. EPA (2017). 40 CFR Part 136, Appendix B: Definition and procedure for the determination of the method detection limit, Revision 2. Legal Information Institute. https://www.law.cornell.edu/cfr/text/40/appendix-B_to_part_136
  8. Johnson, K., Johnson, B., & Johnson, H. (2022). Periodic Table of Elements in the Ocean: summary table (mean concentrations mainly from Nozaki, 1997, Eos 78). MBARI. https://www.mbari.org/know-your-ocean/periodic-table-of-elements-in-the-ocean/summary-table/
  9. Nozaki, Y. (1997). A fresh look at element distributions in the North Pacific. Eos, Transactions AGU. https://doi.org/10.1029/97EO00148
  10. Bruland, K. W., & Lohan, M. C. (2003). Controls of trace metals in seawater. In Treatise on Geochemistry, Vol. 6, 23–47. Elsevier. https://www.whoi.edu/cms/files/6.02_32043.pdf
  11. Bielmyer, G. K., Grosell, M., Bhagooli, R., Baker, A. C., Langdon, C., Gillette, P., & Capo, T. R. (2010). Differential effects of copper on three species of scleractinian corals and their algal symbionts (Symbiodinium spp.). Aquatic Toxicology, 97, 125–133. https://doi.org/10.1016/j.aquatox.2009.12.021
  12. Temerdashev, Z., Abakumov, P., Bolshov, M., Abakumova, D., & Pupyshev, A. (2023). Possibilities and limitations of ICP-spectrometric determination of the total content of tin in waters with different salinity levels, Part 1. Molecules, 28(16), 5967. https://doi.org/10.3390/molecules28165967
  13. Spanos, C. (2016). Analysis of trace elements in seawater using the Thermo Scientific iCAP 7000 Series ICP-OES Duo (Application Note 43185). Thermo Fisher Scientific. https://documents.thermofisher.com/TFS-Assets/CMD/Application-Notes/AN-43185-ICP-OES-Trace-Elements-Seawater-AN43185-EN.pdf
  14. Badocco, D., Di Marco, V., Piovan, A., Caniato, R., & Pastore, P. (2016). A procedure for the quantification of total iodine by inductively coupled plasma mass spectrometry, and its application to the determination of iodine in algae sampled in the lagoon of Venice. Analytical Methods, 8, 7545–7551. https://doi.org/10.1039/C6AY02204A
  15. Millero, F. J. (2010). History of the equation of state of seawater. Oceanography, 23(3), 18–33. https://doi.org/10.5670/oceanog.2010.21
  16. Hanna Instruments (n.d.). Phosphorus and phosphates in your reef tank. Hanna Instruments. https://www.hannainst.com/hubfs/006-finished-content/Aquarium/phosphorus-and-phosphates-in-your-reef-tank–hanna-instruments.pdf
  17. National Research Council Canada (n.d.). NASS-7: Seawater certified reference material for trace metals and other constituents. NRC. https://nrc.canada.ca/en/certifications-evaluations-standards/certified-reference-materials/list/131

Written by AVen for Reef Life. About the author.

Get the next reef guide in your inbox

Science-based articles and tool updates from Reef Life. No spam, unsubscribe anytime.

Read our Privacy Policy.

AVen's SPS reef aquarium

Written by

AVen

Reefkeeper for more than 20 years, with an SPS reef on a deep sand bed. I write science-based guides that start from the research and end in my own tank. Author of the book Reef Dreams.